Elinramstrom (Talk | contribs) |
MatildaBrink (Talk | contribs) |
||
(7 intermediate revisions by 5 users not shown) | |||
Line 14: | Line 14: | ||
#MinION { | #MinION { | ||
− | + | width: 68%; | |
− | + | ||
− | + | ||
− | width: | + | |
} | } | ||
Line 48: | Line 45: | ||
− | <div class="svg-wrapper"> | + | <div class="svg-wrapper" id="Project_Description"> |
Line 209: | Line 206: | ||
<div id="toctitle"></div> | <div id="toctitle"></div> | ||
<ul> | <ul> | ||
− | <li class="toclevel tocsection"><a href="#Project_Description" class="scroll"> <span id="whereYouAre"> | + | <li class="toclevel tocsection"><a href="#Project_Description" class="scroll"> <span id="whereYouAre"> Transciptomics</span> </a> |
<ul> | <ul> | ||
− | <li class="toclevel nav-item active"><a href="# | + | <li class="toclevel nav-item active"><a href="#Seq" class="nav-link scroll"> Sequencing</a></li> |
− | <li class="toclevel nav-item"><a href="# | + | <li class="toclevel nav-item"><a href="#Exp" class="nav-link scroll"> Expermient</a></li> |
− | <li class="toclevel nav-item"><a href="# | + | <li class="toclevel nav-item"><a href="#Res" class="nav-link scroll"> Result</a></li> |
<li class="toclevel nav-item"><a href="#References" class="nav-link scroll"> References </a></li> | <li class="toclevel nav-item"><a href="#References" class="nav-link scroll"> References </a></li> | ||
</ul> | </ul> | ||
Line 245: | Line 242: | ||
<!-- FROM THIS POINT DOWNWARDS YOU START ADDING YOUR STUFF --> | <!-- FROM THIS POINT DOWNWARDS YOU START ADDING YOUR STUFF --> | ||
<div class="card-holder"> | <div class="card-holder"> | ||
− | <h1>Sequencing</h1> | + | <h1 id="Seq">Sequencing</h1> |
</div> | </div> | ||
Line 258: | Line 255: | ||
<p>Sequencing is the final step of the journey and generally makes up a lot of different and varying methods with the common purpose of determining the primary atomic-level structure of what you’re interested in, resulting in sequences of bases making it all up. Most often this includes proteins, polymers or as in our case the keys to life: genetic material; enabling us to read it like a book of sorts.<br><br> | <p>Sequencing is the final step of the journey and generally makes up a lot of different and varying methods with the common purpose of determining the primary atomic-level structure of what you’re interested in, resulting in sequences of bases making it all up. Most often this includes proteins, polymers or as in our case the keys to life: genetic material; enabling us to read it like a book of sorts.<br><br> | ||
− | Oxford Nanopore’s MinION third generation sequencing device | + | Oxford Nanopore’s MinION third generation sequencing device make use of nano-sized holes called nanopores with an applied voltage across them. In essence: when your material of choice passes through one of these pores, each base fluctuates this current by a given amount which acts as a fingerprint to identify the given base [1].<br><br></p> |
Line 268: | Line 265: | ||
<!-- Here goes the big image to the right --> | <!-- Here goes the big image to the right --> | ||
<img src="https://static.igem.org/mediawiki/2018/3/38/T--Uppsala--MinION.jpeg" id="MinION" > | <img src="https://static.igem.org/mediawiki/2018/3/38/T--Uppsala--MinION.jpeg" id="MinION" > | ||
− | <p><b>Figure 1 | + | <p><b>Figure 1.</b> The MinION device.</p> |
</div> | </div> | ||
Line 289: | Line 286: | ||
− | <h2>Experiment</h2> | + | <h2 id="Exp">Experiment</h2> |
<p> Sequencing the final library-prepped RNA includes three primary steps: Priming, loading and sequencing.</p><br> | <p> Sequencing the final library-prepped RNA includes three primary steps: Priming, loading and sequencing.</p><br> | ||
Line 296: | Line 293: | ||
<h3>Loading the Flowcell</h3> | <h3>Loading the Flowcell</h3> | ||
− | <p>Loading the flow cell is a bit tricky, as the wrong action can completely destroy the whole cell. By sucking up the primer mixture carefully with a pipette through a disposal vent, a kind of suction is created over the | + | <p>Loading the flow cell is a bit tricky, as the wrong action can completely destroy the whole cell. By sucking up the primer mixture carefully with a pipette through a disposal vent, a kind of suction is created over the actual pore-membrane area which lets you (with some quick manouvering skills) drop by drop load your sample into that chamber with the help of a pipette which sucks it all down.<br><br> |
Being to slow in this process can result in bubbles forming across the membrane which effectively kills all of the pores. | Being to slow in this process can result in bubbles forming across the membrane which effectively kills all of the pores. | ||
Line 307: | Line 304: | ||
− | <h2>Result</h2> | + | <h2 id="Res">Result</h2> |
<p>Down below are some example pictures from one of our sequencing runs, giving a clue as of what kind of information that can be gathered.</p> | <p>Down below are some example pictures from one of our sequencing runs, giving a clue as of what kind of information that can be gathered.</p> | ||
Line 325: | Line 322: | ||
<div class="inner-card-text"> | <div class="inner-card-text"> | ||
<!-- start of paragraph--> | <!-- start of paragraph--> | ||
− | <p><b>Figure 2 | + | <p><b>Figure 2.</b> State of the sequencing pores after a full sequencing. Notice the amount of sequencing pores (light green) dropping off as issues like saturation (black, orange) start to increase.</p> |
</div> | </div> | ||
<!-- end of paragraph --> | <!-- end of paragraph --> | ||
Line 337: | Line 334: | ||
<div class="inner-card-text"> | <div class="inner-card-text"> | ||
<!-- start of paragraph --> | <!-- start of paragraph --> | ||
− | <p><b>Figure 3 | + | <p><b>Figure 3.</b> Passed versus failed reads.</p> |
<!-- End of paragraphs --> | <!-- End of paragraphs --> | ||
</div> | </div> | ||
Line 370: | Line 367: | ||
<div class="inner-card-text"> | <div class="inner-card-text"> | ||
<!-- start of paragraph--> | <!-- start of paragraph--> | ||
− | <p><b>Figure 4 | + | <p><b>Figure 4.</b> Distribution of read lengths.</p> |
</div> | </div> | ||
<!-- end of paragraph --> | <!-- end of paragraph --> | ||
Line 382: | Line 379: | ||
<div class="inner-card-text"> | <div class="inner-card-text"> | ||
<!-- start of paragraph --> | <!-- start of paragraph --> | ||
− | <p><b>Figure 5 | + | <p><b>Figure 5.</b> Quality score distribution for the different read lengths, were a quality score of at least 7 is acceptable.</p> |
<!-- End of paragraphs --> | <!-- End of paragraphs --> | ||
</div> | </div> | ||
Line 678: | Line 675: | ||
<p>An ideal sequencing run would mean that a lot of the material (depending on run time) has been read and converted into large amounts of ”passed reads” data files containing lines and lines of base sequences. However during our multiple sequencing runs a couple of re-emergin issues consistently showed up, therefore a lot of time was dedicated to troubleshooting this. It was later hypothesized that the source of poor sequencing was because of mRNA left in the samples due to inadequate reverse transcription. See cDNA synthesis </p> | <p>An ideal sequencing run would mean that a lot of the material (depending on run time) has been read and converted into large amounts of ”passed reads” data files containing lines and lines of base sequences. However during our multiple sequencing runs a couple of re-emergin issues consistently showed up, therefore a lot of time was dedicated to troubleshooting this. It was later hypothesized that the source of poor sequencing was because of mRNA left in the samples due to inadequate reverse transcription. See cDNA synthesis </p> | ||
− | <h3>Low | + | <h3>Low Throughput</h3> |
<p>Because it is impossible to have a 100% purely adaptor ligated mRNA sample we initially decided on that it would be acceptable to run the sequencing for a longer time with a below average throughput. However it was noted during repeated sequencing runs that the throughput was actually sub-par, showing that something in the sample(s) was consistently clogging the pores. This meant that only a fraction of our total material was actually being sequenced, eventually leading to near-zero throughputs.</p> | <p>Because it is impossible to have a 100% purely adaptor ligated mRNA sample we initially decided on that it would be acceptable to run the sequencing for a longer time with a below average throughput. However it was noted during repeated sequencing runs that the throughput was actually sub-par, showing that something in the sample(s) was consistently clogging the pores. This meant that only a fraction of our total material was actually being sequenced, eventually leading to near-zero throughputs.</p> | ||
− | <h3>Failed | + | <h3>Failed Reads</h3> |
− | <p>Another reoccuring theme was the issues with ”failed” reads. A failed read consitutes a failed characterization of the base passing through the pores, effectively discarding it. What this means is that whatever came through the pore was actually not cDNA but something foreign.<p | + | <p>Another reoccuring theme was the issues with ”failed” reads. A failed read consitutes a failed characterization of the base passing through the pores, effectively discarding it. What this means is that whatever came through the pore was actually not cDNA but something foreign.<p><br> |
<p>Even though some progress was made creating read-data that passed, the amount of data as well as the issue with consistency in form of duplicate or triplicate sequencing runs made it deemed as not significant enough to carry over to the bioinformatics.</p> | <p>Even though some progress was made creating read-data that passed, the amount of data as well as the issue with consistency in form of duplicate or triplicate sequencing runs made it deemed as not significant enough to carry over to the bioinformatics.</p> | ||
− | |||
− | |||
Line 697: | Line 692: | ||
</div> | </div> | ||
+ | |||
+ | <div class="card-holder"> | ||
+ | <h2 id="References">References</h2> | ||
+ | |||
+ | <p><b>[1]</b> Oxford Nanopore, 2018. How does Nanopore DNA/RNA sequencing work? <a href="https://nanoporetech.com/how-it-works">https://nanoporetech.com/how-it-works</a> Date of visit 2018-10-15</p><br> | ||
+ | </div> | ||
+ | |||
+ | |||
Latest revision as of 16:19, 3 December 2018
Sequencing
Experiment
Sequencing the final library-prepped RNA includes three primary steps: Priming, loading and sequencing.
Priming the Flowcell
The flow cell is essentially where the magic happens. Before the sequencing can take place, a type of storing liquid which protects the flow cell has to be emptied and replaced with a mixture that primes the flow cell, essentially making it ready to sequence. A scan is also performed during this step which tells you how many pores are actually working.
Loading the Flowcell
Loading the flow cell is a bit tricky, as the wrong action can completely destroy the whole cell. By sucking up the primer mixture carefully with a pipette through a disposal vent, a kind of suction is created over the actual pore-membrane area which lets you (with some quick manouvering skills) drop by drop load your sample into that chamber with the help of a pipette which sucks it all down.
Being to slow in this process can result in bubbles forming across the membrane which effectively kills all of the pores.
Sequencing
At this step the sequencing can be started, as the software essentially takes care of everything!
Result
Down below are some example pictures from one of our sequencing runs, giving a clue as of what kind of information that can be gathered.
Figure 2. State of the sequencing pores after a full sequencing. Notice the amount of sequencing pores (light green) dropping off as issues like saturation (black, orange) start to increase.
Figure 3. Passed versus failed reads.
Figure 4. Distribution of read lengths.
Figure 5. Quality score distribution for the different read lengths, were a quality score of at least 7 is acceptable.
Down below are the total of all of our sequencing runs summarised with each run showing runtime, reads, read bases as well as an attached quality score distribution.
Run 1
Run 2
Run 3
Run 4
Run 5
Conclusion
An ideal sequencing run would mean that a lot of the material (depending on run time) has been read and converted into large amounts of ”passed reads” data files containing lines and lines of base sequences. However during our multiple sequencing runs a couple of re-emergin issues consistently showed up, therefore a lot of time was dedicated to troubleshooting this. It was later hypothesized that the source of poor sequencing was because of mRNA left in the samples due to inadequate reverse transcription. See cDNA synthesis
Low Throughput
Because it is impossible to have a 100% purely adaptor ligated mRNA sample we initially decided on that it would be acceptable to run the sequencing for a longer time with a below average throughput. However it was noted during repeated sequencing runs that the throughput was actually sub-par, showing that something in the sample(s) was consistently clogging the pores. This meant that only a fraction of our total material was actually being sequenced, eventually leading to near-zero throughputs.
Failed Reads
Another reoccuring theme was the issues with ”failed” reads. A failed read consitutes a failed characterization of the base passing through the pores, effectively discarding it. What this means is that whatever came through the pore was actually not cDNA but something foreign.
Even though some progress was made creating read-data that passed, the amount of data as well as the issue with consistency in form of duplicate or triplicate sequencing runs made it deemed as not significant enough to carry over to the bioinformatics.
References
[1] Oxford Nanopore, 2018. How does Nanopore DNA/RNA sequencing work? https://nanoporetech.com/how-it-works Date of visit 2018-10-15