Difference between revisions of "Team:TecMonterrey GDL/InterLab"

Line 58: Line 58:
 
The corrected absorbance was was calculated by subtracting the mili Q water reading. The reference OD600 was provided to the team by iGEM. The correction factor was obtained by dividing the reference OD600 by corrected Abs600 obtained.  
 
The corrected absorbance was was calculated by subtracting the mili Q water reading. The reference OD600 was provided to the team by iGEM. The correction factor was obtained by dividing the reference OD600 by corrected Abs600 obtained.  
  
 +
</p>
 +
 +
 +
    <p class= "minisubtitulos">
 +
Calibration Protocol 2: Particle Standard Curve - Microsphere Protocol
 +
    </p>
 +
<p class= "texto1"
 +
Silica microspheres were provided by iGEM kits, these were meant to simulate cells in a dilution, and thus the protocol stated to make serial dilutions of the microsphere suspension in order to construct a standard curve for Abs600 which would allow conversion of Abs600 measurements into an estimated number of cells. Our curve results are shown in the following figure:
 
</p>
 
</p>
 
</div>
 
</div>

Revision as of 02:38, 28 July 2018

InterLab

This year (2018) the iGEM Team TecMonterrey_GDL decided to contribute into the Interlab study measurements. The goal of the study is summarized in achieving a way to reduce lab-to-lab variability in fluorescence measurements via normalizing to absolute cell count or colony forming units. The approaches taken were indicated by the iGEM measurement team: 1. Converting between absorbance of cells to absorbance of a known concentration of microsphere silica beads 2. Counting colony-forming units (CFU) from the samples The team followed a series of protocols created to help standardize the measurement of GFP (green fluorescent protein), which is widely used as measurement marker. Specifications: The equipment we used was Cytation 3 by Biotek. It can measure both absorbance and fluorescence. And the pathlength correction was disabled. The temperature of the measurements was done at room temperature.

Calibration Protocol 1: OD600 Reference point - LUDOX Protocol

The first protocol’s goal was to obtain a conversion factor between absorbance measured and the OD600 measurement as it occurs in spectrophotometry measurements. This was made with the kit provided LUDOX CL-X (45% colloidal silica suspension) provided by iGEM and sterile mili Q water, reactants were added accordingly to the protocol and the obtained results are shown in the next table (Table 1):

como insertar tabla

Table 1. Calibration Protocol 1: OD(600) Reference point - LUDOX Protocol

The corrected absorbance was was calculated by subtracting the mili Q water reading. The reference OD600 was provided to the team by iGEM. The correction factor was obtained by dividing the reference OD600 by corrected Abs600 obtained.

Calibration Protocol 2: Particle Standard Curve - Microsphere Protocol