Line 136: | Line 136: | ||
<p><font size="3"><center>Figure 3. OD<sub>600</sub> (A) and fluorescence (B) interlab study results for positive and negative controls and all test devices after 6 hours. Tests were performed in duplicate using two colonies for each test device, results for which are shown separately to account for colony-to-colony variation. </font></center></p> | <p><font size="3"><center>Figure 3. OD<sub>600</sub> (A) and fluorescence (B) interlab study results for positive and negative controls and all test devices after 6 hours. Tests were performed in duplicate using two colonies for each test device, results for which are shown separately to account for colony-to-colony variation. </font></center></p> | ||
− | <p><font size="3">Results for both 0 and 6 hours were converted into fluorescence per | + | <p><font size="3">Results for both 0 and 6 hours were converted into fluorescence per OD<sub>600</sub> and fluorescence per cell using the previously described calibrations (Figure 4). While the InterLab study is an opportunity for crowdsourcing a large amount of data on the inherent variability of a given biological system, it does not account for variations within data sets. Our team attempted to seek out and address such sources of variation, investigating biodesign automation of protocols, use of internal standards and effects of media composition. Further detail and results are available on the <a href="https://2018.igem.org/Team:Newcastle/Measurement">measurements page</a>.</font></p> |
<img src="https://static.igem.org/mediawiki/2018/a/a5/T--Newcastle--InterlabFigure3.png"> | <img src="https://static.igem.org/mediawiki/2018/a/a5/T--Newcastle--InterlabFigure3.png"> | ||
Line 179: | Line 179: | ||
<div class="col-full"> | <div class="col-full"> | ||
− | + | <p><font size="3">Attributions: Kyle Stanforth, Matthew Burridge, Sam Went</font></p> | |
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | <p | + | |
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
</div> | </div> |
Revision as of 13:28, 16 October 2018
Calibration
Alternative Roots participated in the 2018 Interlab study. Three calibration steps were carried out prior to any experimental measurements being taken. Absorbance and fluorescence values were measured in 96-well plates using a Thermofisher Varioskan Lux plate reader (Thermofisher Scientific). Absorbance was measured at 600 nm and converted to a comparable OD600. Fluorescence was measured at 525 nm with excitation at 485 nm with a 12 nm bandpass width. All readings took place at 25 °C and pathlength correction was disabled. The calibrations were:
- A LUDOX CL-X 45 % colloidal silica suspension was used to calculate a conversion factor for the Abs600 value measured by the plate reader to a comparable OD600 value, considering path length and well volume (table 1). Abs600 of 1:2 dilutions of LUDOX silica suspension were taken in triplicate and a reference OD600 of 0.063 (the reference value for 100 µL of LUDOX CL-X in a well of a standard 96-well flat-bottom black with clear bottom plate) divided by the mean measured value to give a conversion factor.
- A standard curve was prepared by measuring the OD600 of serial dilutions of monodisperse silica microspheres, with similar light scattering properties to E. coli cells. This was used to standardise OD readings across labs (Figure 1 A).
- A fluorescence standard curve was created by measuring the fluorescence of serial dilutions of the small molecule fluorescein. This has similar excitation and emission characteristics to GFP allowing conversion of fluorescence readings to an equivalent fluorescein concentration. Calibrations allowed expression measurement in units of fluorescence per OD and molecules of equivalent fluorescein (MEFL) per cell (Figure 1B).
Protocol
Following these calibration steps, two transformed colonies for each test device and both control plasmids were used to inoculate lysogeny broth (LB) medium containing 25 µg/L of chloramphenicol (CAM) and incubated overnight at 37 °C with shaking at 220 rpm. Overnight cultures were diluted 1:10 and the OD600 adjusted to 0.02 with LB with CAM to a final volume of 12 mL. Fluorescence and Abs600 were taken at 0h and 6 hours of incubation at 37 °C with 220 rpm shaking. Test devices, plasmid backbone and protocol workflow are shown in Figure 2.
Results
Results for OD600 and fluorescence measurements after 6 hours are shown in Figure 3.
Results for both 0 and 6 hours were converted into fluorescence per OD600 and fluorescence per cell using the previously described calibrations (Figure 4). While the InterLab study is an opportunity for crowdsourcing a large amount of data on the inherent variability of a given biological system, it does not account for variations within data sets. Our team attempted to seek out and address such sources of variation, investigating biodesign automation of protocols, use of internal standards and effects of media composition. Further detail and results are available on the measurements page.
InterLab
REFERENCES
Attributions: Kyle Stanforth, Matthew Burridge, Sam Went