Difference between revisions of "Team:Vilnius-Lithuania/InterLab"

Line 93: Line 93:
 
  <p>Using LUDOX CL-X as a single point reference allowed us to obtain a ratiometric conversion factor to transform absorbance data into a standard OD600 measurement. This is crucial to ensure that plate reader measurements are not volume dependent. After this calibration part we obtained a radiometric conversion factor (Tab. 2) which will be used in further Interlab study measurements.</p>
 
  <p>Using LUDOX CL-X as a single point reference allowed us to obtain a ratiometric conversion factor to transform absorbance data into a standard OD600 measurement. This is crucial to ensure that plate reader measurements are not volume dependent. After this calibration part we obtained a radiometric conversion factor (Tab. 2) which will be used in further Interlab study measurements.</p>
  
<p><strong>Tab. 2</strong> LUDOX CL-X measurement. Obtained ratiometric conversion factor is 3,419</p>
+
<strong>Tab. 2</strong> LUDOX CL-X measurement. Obtained ratiometric conversion factor is 3,419
 
<table>
 
<table>
 
     <thead>
 
     <thead>
Line 192: Line 192:
  
  
  <p><strong>Tab. 3</strong> Colony forming units (CFU) per 1 mL of an OD600 = 0.1culture</p>
+
  <strong>Tab. 3</strong> Colony forming units (CFU) per 1 mL of an OD600 = 0.1culture
  
 
  <p><table>
 
  <p><table>
Line 242: Line 242:
 
     </tbody>
 
     </tbody>
 
   </table></p>
 
   </table></p>
 
 
 
 
 
<p></p>
 
  
  

Revision as of 14:06, 17 October 2018

InterLab

Studying Fluorescence

The goal of this year’s InterLab Study was to identify and minimize the sources of systematic variability in fluorescence measurements by normalizing to absolute cell count or colony-forming units (CFUs) instead of optical density (OD).

Participating in the fifth iGEM InterLab Study was a great opportunity to start this year’s competition as well as acquire some valuable knowledge which we implemented into practice during the project.

invert