7/23: > Calibration 1:​ OD​600​ Reference point

Monday, July 23, 2018 9:00

1) Add 100 μl LUDOX into wells A1, B1, C1, D1
2) Add 100 μl of dd H2O into wells A2, B2, C2, D2
3) Measure absorbance at 600 nm of all samples in the measurement mode we plan to use for cell measurements
4) Record the data in both the table below and our notebook
5) Import data into Excel sheet provided

Monday, July 23, 2018 9:25(pm)

Specific data from our experiment


LUDOX CL-X H2O
Replicate 1 0.058 0.038
Replicate 2 0.059 0.037
Replicate 3 0.053 0.037
Replicate 4 0.055 0.036
Arith. Mean 0.056 0.037
Corrected Abs600 0.019
Reference OD600 0.063
OD600/Abs600 3.238