<!DOCTYPE html>
InterLab
We introduced eight plasmids (2 controls and 6 test devices) to E.coli DH5α by following the protocol and then evaluate the measurement by measuring its OD600. We used TECAN infinite M200 PRO as a plate reader.
Transformation
We introduced eight plasmids from the Distribution Kit to E.Coli DH5α.
Calibration
We conducted three tests and calibrated the machine.
Cell Measurement
we measured the OD600 and the fluorescence of the transformed cells.
CFU
We measured the Colony Forming Units using two types of transformation cells.
Transformation
Calibration
Discuss about Calibration
From the Experimental results,Particles / Abs600 = 5.18*10⁸ and MEFL / a.u. = 1.24*10⁹ was found.
Cell Measurement
Discuss about Cell Mesurement
The results show the changes in optical density (λ = 600 nm) of the transformed cells. They show that the cells grew steadily.
However, when we measured the fluorescence, some values appeared negative or go down after 6-hour culturing. Thus, the Fluorescence/OD and MEFL/Particle differed from the theoretical values.
CFU
Discuss abou CFU
We counted the CFU (Colony Forming Unit) and the following figures show the number of colony on each plate. Then, we calculated the CFU / mL in the starting sample when the optical density (λ = 600 nm) is 0.1.