Team:Toronto/InterLab

Calibration 1: OD600 Reference point
  • Purpose of this calibration: To transform absorbance data to a OD600 measurement, calculate a plate-reader specific (Tecan Infinite 200 Pro) conversion factor for OD600 from Abs600 calculated for Ludox CL-X on a mass spectrophotometer.
  • Beer-Lambert’s law of absorbance dictates that optical path length plays a fundamental role in determining absorbance:
  • lambert-law
  • This is necessary because cuvettes (used in photometer) have a fixed path optical path length when using light scattering to measure absorbance as opposed to the varying path lengths of wells in a 96-well plate which change as the volume of sample added in them changes.
  • Results: Cell density readings can thus be converted to OD600 by multiplying correction factor value, 4.138.
Table1.

Table shows absorbance measurements (at 600 nm) for LUDOX CL-X and dd-H2O using a plate reader. The corrected Abs600 is the difference between the LUDOX CL-X reading and dd H2O reading. Reference OD600 is a measurement by a spectrophotometer (provided on iGEM excel sheet). OD600/Abs600 is the correction factor to convert Abs600 to OD600, calculated by dividing Reference OD600 by Abs600.

LUDOX_CL_X Capture Capture1 Capture2 Capture3 Capture4 graph graph1 graph2 graph3 picture picture1 picture2 table2 table3