Development of A Novel
Blood-MicroRNA Handy Detection System with CRISPR
InterLab
Introduction
The InterLab this year has been updated to a more detailed protocol. During the InterLab study, we used the LUDOX, fluorescence and the plasmids that were shipped along with the distribution kit, and closely followed the InterLab study protocol.
Through the InterLab 2018 study, we experienced a lot in the measuring work, and surprisingly got to know more about the instruments that usually were ignored by us. Actually, it is the measuring instruments that help us complete the project every year!!
Methods and Design
1. InterLab Parts
Positive Control (BBa_I20270)
NegativeControl (BBa_R0040)
Test Device1 (BBa_J364000)
Test Device2 (BBa_J364001)
Test Device3 (BBa_J364002)
Test Device4 (BBa_J364007)
Test Device5 (BBa_J364008)
Test Device6 (BBa_J364009)
2. Preparation
To start with, our team transformed E.coli strain DH5α with the provided plasmids, namely Test Device 1,2,3,4,5,6, Positive Control, and Negative Control. As long as colonies had emerged on Cm+ Resistance Media, we picked up 2 colonies in each petrie dish and cultured them at 37℃ with 220 rpm frequency for 14 hours. When the bacteria solutions were turbid enough, we began the following process.
3. OD600 Reference Point
With plate reader, we measured Abs600 of the LUDOX and H2O. The H2O measurement served as the background. Both included 4 technical replicates to enhance the reliability of the results. Comparing to the standard OD600 reference given, which was 0.0425, we were able to achieve a ratio between OD600 and Abs600. The ratio was essential in converting Abs600 raw measurements into standard OD600 records.
4. Fluorescein Standard Curve
Different concentrations of fluorescein were obtained by 2-fold serial dilution. In the first pipet, 200μL Fluorescein 1×stock solution was added. By removing half of the 200μL solution in previous pipets to later ones which already contained 100μL PBS, we were able to generate 11 solutions in half-descending fluorescence concentration. We also included another pipet with 100μL PBS only as blank. Within each concentration, we performed 4 replicates to calculate the more representative mean. After the results were recorded, all fluorescein concentrations were divided by average fluorescence measurement, of which the medium-high mean were calculated to diminish operation errors. This mean point would be used to set up the conversion between fluorescein concentration and fluorescence measurements in later steps.
5. OD600 and Fluorescence Measurements
We tracked the original OD600 of the 16 samples so as to dilute them into 0.02. Once the original values were available, we integrated them into the dilution calculation sheet and followed the suggested volume to dilute each sample. After dilution, we confirmed that OD600 had reached exactly 0.02 or around. Our team set that time as T=0(h) and measured average OD600 from 4 replicates of each 16 samples to reduce technical error. Other 4 replicates were used for T=0(h) fluorescence measurements, with the same equipment and settings in Step 3. Henceforward, we recorded the OD600 and fluorescence when T=0 and T=6.
6. Equipments and Settings
To obtain OD600 measurement, we employed Multiscan FC. Single wavelength of 600nm was used, and path length correction was turned off. To reduce measurement error, we also added a dynamic circulation of 5 times, with 2-second intervals.
The fluorescence measurement was obtained through Fluoroskan Ascent FL and Thermo Fisher. The filters used were 485nm and 538nm for excitation and emission respectively. The measurement cycled 5 times with intervals of 2 second.
Results & Discussion
Calibration
Calibration1: OD600 Reference point – LUDOX Protocol
We used LUDOX CL-X (45% colloidal silica suspension) , ddH2O and 96 well plate(black with clear flat bottom preferred) to get the conversion factor, which can transform Abs600 measurements into comparable OD600 measurements.
LUDOX CL-X | H2O | |
---|---|---|
Replicate 1 | 0.062 | 0.039 |
Replicate 2 | 0.060 | 0.039 |
Replicate 3 | 0.066 | 0.038 |
Replicate 4 | 0.061 | 0.038 |
Arith. Mean | 0.062 | 0.039 |
Corrected Abs600 | 0.024 | |
Reference OD600 | 0.063 | |
OD600/Abs600 | 2.654 |
Calibration2: Particle Standard Curve – Microsphere Protocol
Following the protocol, we prepared a dilution series of monodisperse silica microspheres and measure the Abs600 in our plate reader. We got a particle standard curve and transferred it into log scale, which are shown as below.
Fig.2 Particle standard curve
This is the sign of the famous game!
Calibration3: Fluorescence standard curve – Fluorescein Protocol
We did a dilution series of fluorescein in four replicates. After recording the measurements, we generated a standard curve of fluorescence for fluorescein concentration and its log scale version, which are shown as below.
Calibration
hello everyone this is the another example-content part for more imformation. Hello everyone this is the another example-content part for more imformation.Donec id elit non mi porta gravida at eget metus. Fusce dapibus, tellus ac cursus commodo, tortor mauris condimentum nibh, ut fermentum massa justo sit amet risus. Etiam porta sem malesuada magna mollis euismod. Donec sed odio dui.
Donec id elit non mi porta gravida at eget metus. Fusce dapibus, tellus ac cursus commodo, tortor mauris condimentum nibh, ut fermentum massa justo sit amet risus. Etiam porta sem malesuada magna mollis euismod. Donec sed odio dui.
Maybe you want to write something interesting thing here! Someone famous in Source Title
hello everyone this is the another example-content part for more imformation. Hello everyone this is the another example-content part for more imformation.Donec id elit non mi porta gravida at eget metus. Fusce dapibus, tellus ac cursus commodo, tortor mauris condimentum nibh, ut fermentum massa justo sit amet risus. Etiam porta sem malesuada magna mollis euismod. Donec sed odio dui.
Donec id elit non mi porta gravida at eget metus. Fusce dapibus, tellus ac cursus commodo, tortor mauris condimentum nibh, ut fermentum massa justo sit amet risus. Etiam porta sem malesuada magna mollis euismod. Donec sed odio dui.
This is the sign of the famous game!
hello everyone this is the another example-content part for more imformation. Hello everyone this is the another example-content part for more imformation.Donec id elit non mi porta gravida at eget metus. Fusce dapibus, tellus ac cursus commodo, tortor mauris condimentum nibh, ut fermentum massa justo sit amet risus. Etiam porta sem malesuada magna mollis euismod. Donec sed odio dui.
Donec id elit non mi porta gravida at eget metus. Fusce dapibus, tellus ac cursus commodo, tortor mauris condimentum nibh, ut fermentum massa justo sit amet risus. Etiam porta sem malesuada magna mollis euismod. Donec sed odio dui.
LUDOX CL-X | H2O | |
---|---|---|
Replicate 1 | 0.062 | 0.039 |
Replicate 2 | 0.060 | 0.039 |
Replicate 3 | 0.066 | 0.038 |
Replicate 4 | 0.061 | 0.038 |
Arith. Mean | 0.062 | 0.039 |
Corrected Abs600 | 0.024 | |
Reference OD600 | 0.063 | |
OD600/Abs600 | 2.654 |
Example block-level help text here.There are several variants of derived from different organisms. In our project, we . The structure of this protein has been determined with and without The a lobe. The two lobes are positively charged towards the protein core to accommodate the negatively charged RNA. Each of these two lobes contains an RNase domain. At the main is responsible for target cleavage. In contrast to other two RNase domains of the NUC lobe are located at the outside of the protein, which is likely the reason collateral cleavage upon activation by binding to a matching target. These two domains have been labeled as red spots in Figure 2 and can be found at the interface between the green and pink domain.