Team:CO Mines/InterLab

For the 2018 iGem Competition, we participated in the iGem Interlab Study, which seeks to reduce variability in fluorescence measurements. The Interlab Protocol provided by iGEM was followed.

The InterLab gave us a lot of experience with procedures that we ended up using later. Our project required florescence measurements, so practicing with the 96 well plate reader beforehand was beneficial. InterLab also gave us practice with serial dilutions, which was used to create the cadmium ranges for the kinetic tests.

There were no notable errors when following the iGEM procedure. The process did take longer than we originally expected, but everything worked out as planned. The plate counting might have contained not have been accurate due to human error, but the data seems consistent with the expected results.

With the InterLab procedure, we seemed to successfully calibrate the plate reader for the absorbance and fluorescence. The different devices showed different amounts of fluorescence with similar absorbance. The raw data and some of the resulting graphs are shown in below.

For best display of data, please maximize this window or view in fullscreen on a laptop.

Calibration 1: OD600 Reference Point

Silica Bead Standardization

LUDOX CL-X H2O
Replicate 1 0.056 0.036
Replicate 2 0.055 0.036
Replicate 3 0.065 0.036
Replicate 4 0.062 0.036
Arith. Mean 0.060 0.036
Corrected Abs600 0.024
Reference OD600 0.063
OD600/Abs600 2.681

Table 1: OD600 calibration results, which are used to standardize the results against the Ludox Cl-X beads.

Calibration 2: Particle Standard Curve

The Abs600 of a serial dilution of the silica beads was measured to determine the absorbance over particle number of the beads. This is used to estimate the relationship between the absorbance of a culture of E. coli and the number of cells since the beads have a similar size and shape to E. coli.

Number of Particles 2.35E8 1.18E8 5.88E7 2.94E7 1.47E7 7.35E6 3.68E6 1.84E6 9.19E5 4.60E5 2.30E5 0
Replicate 1 0.684 0.375 0.225 0.155 0.117 0.102 0.093 0.089 0.082 0.082 0.083 0.083
Replicate 2 0.941 0.495 0.239 0.166 0.120 0.103 0.093 0.087 0.082 0.082 0.083 0.082
Replicate 3 0.812 0.629 0.181 0.158 0.120 0.103 0.094 0.085 0.085 0.083 0.083 0.085
Replicate 4 0.797 0.219 0.125 0.222 0.119 0.105 0.090 0.085 0.085 0.084 0.082 0.082
Arith. Mean 0.809 0.430 0.193 0.175 0.119 0.103 0.093 0.087 0.084 0.083 0.083 0.083
Arith. Std. Dev. 0.105 0.175 0.051 0.032 0.001 0.001 0.002 0.002 0.002 0.001 0.001 0.001
Arith. Net Mean 0.726 0.347 0.110 0.092 0.036 0.020 0.009 0.004 0.001 0.000 0.000

Table 2: 'Particle Standard Curve' raw data.

Figure 1: 'Particle Standard Curve' comparing absorbance to particle count.

Figure 2: 'Particle Standard Curve' comparing absorbance to particle count, logarithmic scale.

Number of Particles 2.35E8 1.18E8 5.88E7 2.94E7 1.47E7 7.35E6 3.68E6 1.84E6 9.19E5 4.60E5 2.30E5
Mean particles / Abs600 3.24E8 3.40E8 5.37E8 3.19E8 4.08E8 3.63E8 3.87E8 5.25E8 1.84E9 1.84E9 9.19E8
Mean of med-high levels: 3.93E8

Table 3: 'Particle Standard Curve' number of bead particles per Abs600.

Calibration 3: Fluorescence Standard Curve

Fluorescein has similar fluorescence to GFP, so it is used to normalize GFP concentration to fluorescence at varying concentrations of fluorescein.

Fluorescein uM 10 5 2.5 1.25 0.625 0.313 0.156 0.078 0.039 0.0195 0.0098 0
Replicate 1 8.549E6 4.618E6 2.339E6 1.530E6 4.857E5 2.274E5 1.217E5 2.427E4 2.269E4 1.072E4 5.200E3 9.455E2
Replicate 2 8.730E6 4.573E6 2.556E6 1.363E6 4.682E5 2.189E5 1.231E5 4.901E4 2.978E4 1.229E4 6.461E3 0
Replicate 3 9.203E6 5.201E6 1.724E6 7.605E5 3.571E5 2.384E5 9.723E4 5.153E4 2.884E4 1.056E4 6.146E3 1.576E2
Replicate 4 8.661E6 4.541E6 2.354E6 1.123E6 5.169E5 2.266E5 9.455E4 6.697E4 2.584E4 9.613E3 4.570E3 9.455E2
Arith. Mean 8.786E6 4.733E6 2.243E6 1.194E6 4.570E5 2.278E5 1.091E5 4.795E4 2.679E4 1.079E4 5.594E3 5.122E2
Arith. Std. Dev. 2.883E5 3.134E5 3.602E5 3.340E5 6.956E4 8.041E3 1.534E4 1.767E4 3.206E3 1.111E3 8.679E2 5.045E2
Arith. Net Mean 8.785E6 4.733E6 2.243E6 1.194E6 4.564E5 2.273E5 1.086E5 4.743E4 2.628E4 1.028E4 5.082E3 0

Table 4: 'Fluorescein Standard Curve' raw data.

Figure 3: 'Fluorescence Standard Curve' comparing fluorescence to concentration.

Figure 4: 'Fluorescence Standard Curve' comparing fluorescence to concentration, logarithmic scale.

Fluorescein uM 10 5 2.5 1.25 0.625 0.313 0.156 0.078 0.039 0.0195 0.0098
uM Fluorescein / a.u. 1.14E-6 1.06E-6 1.11E-6 1.05E-6 1.37E-6 1.37E-6 1.44E-6 1.65E-6 1.49E-6 1.90E-6 1.92E-6
Mean uM fluorescein / a.u. 1.190E-6
MEFL / a.u. 7.180E+06

Table 5: 'Fluorescein Standard Curve' relative fluorescein ratios.

Cell Measurement

DH5-alpha E. coli were transformed with the GFP inducible plasmids provided in the iGEM test kit according to the recommended transformation protocol.

Table 6: Fluorescence raw data.

Table 7: Abs600 raw data.

Table 8: Calculated fluorescence over OD. (uM Fluorescein / OD)

Table 9: Calculated fluorescence per cell. (MEFL / particle)

Determining Colony Forming Units

Colony Replicate Dilute 1 Dilute 2 Dilute 3
1 1 1290 890 355
1 2 595 653 238
1 3 1076 619 123
2 1 1160 620 242
2 2 1262 352 128
2 3 1027 903 215
3 1 369 102 43
3 2 302 131 29
3 3 289 152 81
4 1 506 921 183
4 2 831 169 103
4 3 812 796 236

Table 10: Colonies per plate of transformed E. coli.

Colony Replicate Dilute 1 Dilute 2 Dilute 3
1 1 1.03E8 7.12E8 2.84E9
1 2 4.76E7 5.22E8 1.90E9
1 3 8.61E7 4.95E8 9.84E8
2 1 9.28E7 4.96E8 1.94E9
2 2 1.01E8 2.82E8 1.02E9
2 3 8.22E7 7.22E8 1.72E9
3 1 2.95E7 8.16E7 3.44E8
3 2 2.42E7 1.05E8 2.32E8
3 3 2.31E7 1.22E8 6.48E8
4 1 4.05E7 7.37E8 1.46E9
4 2 6.65E7 1.35E8 8.24E8
4 3 6.50E7 6.37E8 1.89E9

Table 11: Colony forming units, calculated from dilutions of plated cultures.

The number of cells in Table 11 can be normalized to the absorbance of the beads in Table 2 so that a general comparison of the absorbance of a culture can be expressed in terms of bead concentration. The fluorescence of the cultures in Table 9 can be compared to the fluorescence of fluorescein molecules to standardize the arbitrary fluorescence measurements commonly used in biology.