Oscarliu117 (Talk | contribs) |
|||
Line 56: | Line 56: | ||
<ul> | <ul> | ||
− | <li>Because the analog value returned by each sensor may be inaccurate, and | + | <li class="none">Because the analog value returned by each sensor may be inaccurate, and |
the carbon dioxide concentration in each area may be slightly | the carbon dioxide concentration in each area may be slightly | ||
different, it must be corrected before use.</li></br> | different, it must be corrected before use.</li></br> | ||
Line 110: | Line 110: | ||
<ul> | <ul> | ||
− | <li>We can then get the unknown CO<sub>2</sub> concentration from this formula.</li> | + | <li class="none">We can then get the unknown CO<sub>2</sub> concentration from this formula.</li> |
</ul> | </ul> | ||
Line 116: | Line 116: | ||
<ul> | <ul> | ||
− | <li>We put CO<sub>2</sub> sensor into the CO<sub>2</sub> incubator and measure its analog value.</li> | + | <li class="none">We put CO<sub>2</sub> sensor into the CO<sub>2</sub> incubator and measure its analog value.</li> |
</ul> | </ul> | ||
Line 131: | Line 131: | ||
<li>Purpose</li></br> | <li>Purpose</li></br> | ||
<ul> | <ul> | ||
− | <li>To verify whether the logarithmic concentration of carbon dioxide is | + | <li class="none">To verify whether the logarithmic concentration of carbon dioxide is |
negatively linear with the output analog value.</li></br> | negatively linear with the output analog value.</li></br> | ||
</ul> | </ul> | ||
Line 138: | Line 138: | ||
<ul> | <ul> | ||
− | <li>Use a CO<sub>2</sub> incubator to adjust the known carbon dioxide concentration and measure its analogy to see if it is on this line.</li></br> | + | <li class="none">Use a CO<sub>2</sub> incubator to adjust the known carbon dioxide concentration and measure its analogy to see if it is on this line.</li></br> |
</ul> | </ul> | ||
Line 145: | Line 145: | ||
<ul> | <ul> | ||
− | <li>We put our sensor in the environment of CO<sub>2</sub> 2.7%, and we measured that | + | <li class="none">We put our sensor in the environment of CO<sub>2</sub> 2.7%, and we measured that |
the analog value 550, which is quite match with the theoretical analog | the analog value 550, which is quite match with the theoretical analog | ||
value 552.</li></br> | value 552.</li></br> | ||
Line 156: | Line 156: | ||
<ul> | <ul> | ||
− | <li>It is best to preheat for the first time for 24 hours, use it for more than | + | <li class="none">It is best to preheat for the first time for 24 hours, use it for more than |
six hours, preheat for 1-2 hours, power off for more than 72 hours, and | six hours, preheat for 1-2 hours, power off for more than 72 hours, and | ||
preheat for 24 hours.</li> | preheat for 24 hours.</li> |
Revision as of 13:52, 17 October 2018
CO2 Sensor