Difference between revisions of "Team:NCKU Tainan/pH meter"

Line 62: Line 62:
  
 
                                         <ul>
 
                                         <ul>
                                             <li>Because the buffer solution wouldn’t change too much in the
+
                                             <li class="none">Because the buffer solution wouldn’t change too much in the
 
                                                 environment, we used the buffer solution to calibrate the pH meter to
 
                                                 environment, we used the buffer solution to calibrate the pH meter to
 
                                                 ensure that the deviation would be the same every time.</li></br>
 
                                                 ensure that the deviation would be the same every time.</li></br>
Line 133: Line 133:
 
                                         <li>Purpose</li></br>
 
                                         <li>Purpose</li></br>
 
                                         <ul>
 
                                         <ul>
                                             <li>We assume that pH 4 to 7 is linear, so we want to verify
+
                                             <li class="none">We assume that pH 4 to 7 is linear, so we want to verify
 
                                                 whether it is linear between pH
 
                                                 whether it is linear between pH
 
                                                 4 to 7. By measuring the solution of pH 4.7, compare its
 
                                                 4 to 7. By measuring the solution of pH 4.7, compare its
Line 142: Line 142:
  
 
                                         <ul>
 
                                         <ul>
                                             <li>Mix a solution with pH=4.7 and measure with a calibrated pH
+
                                             <li class="none">Mix a solution with pH=4.7 and measure with a calibrated pH
 
                                                 meter.</li></br>
 
                                                 meter.</li></br>
 
                                         </ul>
 
                                         </ul>
Line 150: Line 150:
  
 
                                         <ul>
 
                                         <ul>
                                             <li>After inserting the solution with pH=4.7, we found that the
+
                                             <li class="none">After inserting the solution with pH=4.7, we found that the
 
                                                 value was stable at pH 4.83
 
                                                 value was stable at pH 4.83
 
                                                 with an error of about 0.13, which roughly met the error of
 
                                                 with an error of about 0.13, which roughly met the error of

Revision as of 13:48, 17 October 2018