<li>Experiment 1:Instrument calibration</li> | <li>Experiment 1:Instrument calibration</li> | ||
− | |||
− | |||
− | |||
− | |||
− | |||
− | <div class="carousel-item active" style="background-image: url('http://placehold.it/1900x1080')"></div> | + | <ol><li>Experimental method</li> |
+ | |||
+ | <ul><li>(1)Insert pH meter into pH 7 buffer solution, wait about 1 min, it will achieve a stable value.</li> | ||
+ | <li>(2)Minus the value with pH 7, and it will get the offset value. For instance, 7-7.09=-0.09.</li> | ||
+ | |||
+ | <div class="carousel-item active" style="background-image: url('http://placehold.it/1900x1080')"></div> | ||
− | + | <li>(3)Write the offset value into code, upload the code into Arduino again.</li> | |
− | + | <div class="carousel-item active" style="background-image: url('http://placehold.it/1900x1080')"></div> | |
− | + | <li>(4) After rinsing the pH meter, insert it into pH 4 buffer solution.</li> | |
− | + | <li>(5) If it is found to be too different from the error of pH=4, you should adjust the knob. After inserting pH = 4, it is found that the error with 4 is very large.</li> | |
− | + | ||
+ | <div class="carousel-item active" style="background-image: url('http://placehold.it/1900x1080')"></div> | ||
+ | |||
+ | <li>Adjust the knob. (Remark: Because the change of the knob is small, it may have to turn a few more times until display pH 4.)</li> | ||
− | + | <div class="carousel-item active" style="background-image: url('http://placehold.it/1900x1080')"></div> | |
− | + | <li>Finally, it can be found that the measured pH is 4.</li> | |
− | + | ||
− | + | ||
− | + | <div class="carousel-item active" style="background-image: url('http://placehold.it/1900x1080')"></div></ul></ul> | |
− | |||
<p class="pcontent">2. Experimental formula calculation</br> | <p class="pcontent">2. Experimental formula calculation</br> |