− | <p class=" | + | <p class="f12">Synthetic biology interconnects the principles of biology and engineering. Reliable and repeatable measurement is one of the fundamentals of engineering, which synthetic biology must also aspire to achieve. However, replicability of fluorescence measurements between labs presents a unique challenge because of differences in how the data are acquired and processed, and other sources of systematic variability that are not accounted for (such as the number of cells in a sample). By using “relative expression” comparisons, users from different labs cannot directly compare results. <br><br> |
The purpose of the 2018 InterLab Study was to attempt to reduce lab-to-lab variability by standardizing green fluorescent protein (GFP) fluorescence measurements. This was done by normalizing GFP fluorescence to absolute cell count or colony-forming units (CFUs) instead of optical density (OD).</p> | The purpose of the 2018 InterLab Study was to attempt to reduce lab-to-lab variability by standardizing green fluorescent protein (GFP) fluorescence measurements. This was done by normalizing GFP fluorescence to absolute cell count or colony-forming units (CFUs) instead of optical density (OD).</p> | ||
</div> | </div> | ||
Line 29: | Line 29: | ||
<h3>Overview</h3> | <h3>Overview</h3> | ||
− | <p class=" | + | <p class="f12"> Two approaches were used:<br><br> |
<b>1. Converting between absorbance of cells to absorbance of a known concentration of beads.</b><br> | <b>1. Converting between absorbance of cells to absorbance of a known concentration of beads.</b><br> | ||
Line 40: | Line 40: | ||
<h3>Our Equipment</h3> | <h3>Our Equipment</h3> | ||
− | <p class=" | + | <p class="f12">A Molecular Devices Spectramax i3x plate reader was used to measure sample absorbance and fluorescence. This device has variable temperature settings, top optics, and pathlength correction, which can be disabled. All GFP measurements were taken at wavelengths of 535/25nm for emission and 485/20nm for excitation. We used clear-bottomed 96-well plates for all measurements.</p><br><br> |
<h3> Calibration </h3> | <h3> Calibration </h3> | ||
− | <p class=" | + | <p class="f12">1. LUDOX CL-X (45% colloidal silica suspension) was used as a single point reference to obtain a |
conversion factor to transform absorbance data into a comparable OD measurement.<br><br> | conversion factor to transform absorbance data into a comparable OD measurement.<br><br> | ||
Line 55: | Line 55: | ||
<h3> Cell Measurement Protocol </h3> | <h3> Cell Measurement Protocol </h3> | ||
− | <p class=" | + | <p class="f12">Briefly, we measured the raw absorbance and fluorescence of diluted cells (transformed with the below testing devices) at 0 hours and after 6 hours of incubation at 37°C.</p><br></br> |
Line 121: | Line 121: | ||
<h1>Results</h1> | <h1>Results</h1> | ||
− | <p class=" | + | <p class="f12">In our standard curves, we expected a linear relationship between concentration and fluorescence/absorbance. However, the log-scale particle standard curve (Figure 2B) does not exhibit a linear relationship. Possible sources of error include pipetting error or oversaturation of the detector.</p><br><br> |
</div> | </div> | ||
</div> | </div> |
Revision as of 19:54, 17 October 2018
INTERLAB STUDY
Synthetic biology interconnects the principles of biology and engineering. Reliable and repeatable measurement is one of the fundamentals of engineering, which synthetic biology must also aspire to achieve. However, replicability of fluorescence measurements between labs presents a unique challenge because of differences in how the data are acquired and processed, and other sources of systematic variability that are not accounted for (such as the number of cells in a sample). By using “relative expression” comparisons, users from different labs cannot directly compare results.
The purpose of the 2018 InterLab Study was to attempt to reduce lab-to-lab variability by standardizing green fluorescent protein (GFP) fluorescence measurements. This was done by normalizing GFP fluorescence to absolute cell count or colony-forming units (CFUs) instead of optical density (OD).
Methods
Overview
Two approaches were used:
1. Converting between absorbance of cells to absorbance of a known concentration of beads.
Absorbance is a measure of how light is scattered by cells in liquid, which can be used to estimate the concentration of cells in the sample. Assuming that silica beads approximately the same size and shape of E. coli cells would have similar optical properties, we can use the absorbance measurement of a known concentration of beads to produce a standardized “equivalent concentration of beads” measurement.
2. Counting colony-forming units (CFUs) from the sample.
We can estimate the number of individual cells in a volume of liquid media by plating a sample and quantifying colonies (which are each clones of a single cell). The number of colonies in positive and negative control samples can be used to calculate a conversion factor from absorbance to CFU.
For complete methods, refer to: https://2018.igem.org/Measurement/InterLab/Plate_Reader
Our Equipment
A Molecular Devices Spectramax i3x plate reader was used to measure sample absorbance and fluorescence. This device has variable temperature settings, top optics, and pathlength correction, which can be disabled. All GFP measurements were taken at wavelengths of 535/25nm for emission and 485/20nm for excitation. We used clear-bottomed 96-well plates for all measurements.
Calibration
1. LUDOX CL-X (45% colloidal silica suspension) was used as a single point reference to obtain a
conversion factor to transform absorbance data into a comparable OD measurement.
2. A dilution series of monodisperse silica microspheres was used to measure absorbance and produce a standard curve of particle concentration. This standard curve was later used to convert absorbance measurements to an estimated number of cells.
3. A dilution series of fluorescein, which has similar excitation and emission properties to GFP, was used to generate a fluorescein standard curve. This standard curve was later used to convert cell-based readings to an equivalent fluorescein concentration.
Cell Measurement Protocol
Briefly, we measured the raw absorbance and fluorescence of diluted cells (transformed with the below testing devices) at 0 hours and after 6 hours of incubation at 37°C.
Device | Part number | Plate | Location |
---|---|---|---|
Negative Control | BBa_R0040 | Kit Plate 7 | Well 2D |
Positive Control | BBa_I20270 | Kit Plate 7 | Well 2B |
Test Device 1 | BBa_J364000 | Kit Plate 7 | Well 2F |
Test Device 2 | BBa_J364001 | Kit Plate 7 | Well 2H |
Test Device 3 | BBa_J364002 | Kit Plate 7 | Well 2J |
Test Device 4 | BBa_J364007 | Kit Plate 7 | Well 2L |
Test Device 5 | BBa_J364008 | Kit Plate 7 | Well 2N |
Test Device 6 | BBa_J364009 | Kit Plate 7 | Well 2P |
Results
In our standard curves, we expected a linear relationship between concentration and fluorescence/absorbance. However, the log-scale particle standard curve (Figure 2B) does not exhibit a linear relationship. Possible sources of error include pipetting error or oversaturation of the detector.
Figure 1: A) Fluorescein standard curve. B) Fluorescein standard curve (log-scale).
Figure 2: A) Particle standard curve. B) Particle standard curve (log-scale).
Figure 3: A) Average fluorescence per optical density. B) Average fluorescence per particle.