In order to attain bronze medal requirement, IGEM teams have to participate in the InterLab study. Laboratories from around the world conducted these experiments. In this manner, their studies are trustworthy and repeatable, the key component of scientific discipline. Thus, other related labs will be able to build upon others’ work.

Indeed, fluorescence is a property commonly measured in synthetic biology, but data from different labs are not comparable due to different units have been used. The data has been processed differently and the outcome is given in ‘relative expression’. The InterLab study aims to counter this issue by using detailed protocol and data analysis to yield absolute units for measuring GFP. This year, we are focusing on reducing variable fluorescence measurements between different labs by normalizing to absolute cell count or colony-forming units (CFUs) instead of OD.

Experiments

❏ Instrument brand and model _BMG FLUO star OPTIMA__ ❏ Can your instrument measure both absorbance and fluorescence? _YES____ ❏ Does your instrument have pathlength correction, and if yes can it be disabled? _YES; YES_________________________________________________________ ❏ Does your instrument have variable temperature settings, and if yes can this be set to room temperature? YES; YES ❏ What filters does your instrument have for measuring GFP? You will need information about the bandpass width (530 nm / 30 nm bandpass, 25-30nm width is recommended), excitation (485 nm is recommended) and emission (520-530 nm is recommended) of this filter. 485nm; 520nm ❏ Does your instrument use top or bottom optics (i.e. does your plate reader read samples from the top of the plate or the bottom)? _Can read from both direction

Results

Calibration

OD600 reference point

LUDOX CL-XH2O
Replicate 10.1720.124
Replicate 20.1850.106
Replicate 30.1760.137
Replicate 40.1720.124
Arith. Mean0.1760.123
Corrected Abs6000.054 
Reference OD6000.063 
OD600/Abs6001.178

Particle standard curve

Number of Particles2.35E+081.18E+085.88E+072.94E+071.47E+077.35E+063.68E+061.84E+069.19E+054.60E+052.30E+050
Replicate 11.0170.6030.3940.2840.2260.1960.1820.1720.1570.1640.1640.098
Replicate 21.0120.6230.3980.2890.2440.2120.190.1790.1760.1730.1720.101
Replicate 30.990.620.3990.2880.2420.2060.180.1750.1730.1640.1640.107
Replicate 41.0930.570.3770.280.230.20.1830.1860.1740.1760.1680.123
Arith. Mean1.0280.6040.3920.2850.2360.2040.1840.1780.1700.1690.1670.107
Arith. Std.Dev.0.0450.0240.0100.0040.0090.0070.0040.0060.0090.0060.0040.011
Arith. Net Mean0.9210.4970.2850.1780.1280.0960.0770.0710.0630.0620.060

Fluorescein standard curve

Fluorescein uM10.0052.51.250.6250.3130.1560.0780.0390.01950.00980
Replicate 12.327E+041.229E+046.462E+033.505E+031.802E+039.030E+024.610E+022.390E+021.240E+026.100E+013.100E+011.000E+00
Replicate 22.402E+041.211E+046.585E+033.380E+031.821E+038.900E+024.860E+022.420E+021.160E+026.800E+013.300E+011.000E+00
Replicate 32.405E+041.224E+046.671E+033.334E+031.739E+038.830E+024.690E+022.490E+021.230E+026.200E+013.800E+012.000E+00
Replicate 42.301E+041.231E+046.560E+033.403E+031.709E+038.800E+024.520E+022.380E+021.210E+026.100E+013.000E+011.000E+00
Arith. Mean2.359E+041.224E+046.570E+033.406E+031.768E+038.890E+024.670E+022.420E+021.210E+026.300E+013.300E+011.250E+00
Arith. Std.Dev.5.267E+029.108E+018.600E+017.227E+015.256E+011.023E+011.445E+014.967E+003.559E+003.367E+003.559E+005.000E-01
Arith. Net Mean2.359E+041.224E+046.568E+033.404E+031.767E+038.878E+024.658E+022.408E+021.198E+026.175E+013.175E+01

Cell Measurement

Fluorescence Raw Readings:

Hour 0:Neg. ControlPos. ControlDevice 1Device 2Device 3Device 4Device 5Device 6LB + Chlor (blank)
Colony 1, Replicate 1728212599701371308470
Colony 1, Replicate 2678512497701341278273
Colony 1, Replicate 36984130102671141248270
Colony 1, Replicate 46984130102671141248270
Colony 2, Replicate 17781118103641411218268
Colony 2, Replicate 26781118102721371218168
Colony 2, Replicate 37383120101761361228568
Colony 2, Replicate 47085121104701231216773

Hour 6:Neg. ControlPos. ControlDevice 1Device 2Device 3Device 4Device 5Device 6LB + Chlor (blank)
Colony 1, Replicate 18313702012086123260180124061
Colony 1, Replicate 28313671912166123240178121166
Colony 1, Replicate 39013121982154120230173119863
Colony 1, Replicate 48810571992074120193170121761
Colony 2, Replicate 19314221382231118440160119163
Colony 2, Replicate 28613021272195123399152118362
Colony 2, Replicate 38513361262303120414150116959
Colony 2, Replicate 48612031212195116399150116961

Abs600 Raw Readings:

Hour 0:Neg. ControlPos. ControlDevice 1Device 2Device 3Device 4Device 5Device 6LB + Chlor (blank)
Colony 1, Replicate 10.1140.1060.1090.1060.1090.1040.1040.1030.109
Colony 1, Replicate 20.1210.1160.1140.1140.1130.110.1120.1080.112
Colony 1, Replicate 30.1190.1120.1150.1170.1140.1120.1120.1110.112
Colony 1, Replicate 40.1270.1210.1240.1190.1240.1220.120.1340.115
Colony 2, Replicate 10.1210.1160.1160.1160.120.1150.1160.120.123
Colony 2, Replicate 20.1120.1210.120.1190.1260.1150.1190.1160.118
Colony 2, Replicate 30.1160.1130.1190.1180.1170.1140.1120.1130.114
Colony 2, Replicate 40.1230.1180.1180.120.120.1150.1180.1140.117

Hour 6:Neg. ControlPos. ControlDevice 1Device 2Device 3Device 4Device 5Device 6LB + Chlor (blank)
Colony 1, Replicate 10.820.790.1440.7560.8230.1220.1280.8170.11
Colony 1, Replicate 20.8090.8350.1550.7980.8290.1260.1350.8090.111
Colony 1, Replicate 30.840.7940.1920.7840.8290.1490.1340.810.127
Colony 1, Replicate 40.8620.660.1550.7720.8660.1210.1580.8110.153
Colony 2, Replicate 10.8810.830.1240.8020.8340.1710.120.8320.135
Colony 2, Replicate 20.8630.8030.1250.8220.8580.1870.1280.830.128
Colony 2, Replicate 30.850.8170.1390.8240.8330.1870.1420.8250.142
Colony 2, Replicate 40.8630.7390.1480.8120.8350.1880.1470.8250.143

Colony-Forming Unit

p1  p2  n1  n2  
D3_ __282________
D412415413591628011695120116184103
D521401712782013201012310