<a style="top: 0%; left: 50%; width: 50%; height: 100%;"><img src="https://static.igem.org/mediawiki/2018/d/d9/T--UChile_Biotec--4int.png" style="width:100%;object-fit:cover;" id="elem"/></a> | <a style="top: 0%; left: 50%; width: 50%; height: 100%;"><img src="https://static.igem.org/mediawiki/2018/d/d9/T--UChile_Biotec--4int.png" style="width:100%;object-fit:cover;" id="elem"/></a> | ||
</div> | </div> | ||
− | < | + | <center>Figure 2: Fluorescein Standard curves with their respective linear equations. (See normal scale left and log scale right).</center> |
</center> | </center> | ||
− | + | Once done the calibrations we moved to the next steps in the Interlab measurement. | |
</li> | </li> |
Revision as of 07:24, 17 October 2018
InterLab
Calibrations
Three calibrations were performed prior to have standard and more reliable results of Optical Density and Fluorescence in the Cell Measurement. The calibrations performed are next:
- LUDOX: We used LUDOX CL-X (45% colloidal silica suspension) to transform the absorbance (Abs600) to a corrected value of optical density (OD600) taking in consideration the specifications of the plate. The results of this calibration can be seen in Tables I and II.
Table I: Abs600 raw values obtained in the measurements. Table II: Treated values based on the data in Table I. - Microsphere Beads: We used monodisperse silica microspheres solution to measure Abs600, the principal idea of this calibration is to standardize a calibration curve of microspheres concentration to be used to convert Abs600 measurements (see Cell Measurements) to an estimated number of cells. We used microsphere beads because those have a similar size and optical characteristics of the cells used in the Cell Measurements. The results of this calibration can be seen in Figure 1.
Figure 1: Particle Standard curves with their respective linear equations. (See normal scale left and log scale right). - Fluorescein: We used Fluorescein to make a fluorescence curve, fortunately Fluorescein have similar excitation and emission properties than GFP (the molecule that we measured in the Cell Measurement), and is perfect to make a cost-effective calibration curve of fluorescence to know the amount of GFP in the cells we measured. The results of this calibration can be seen in Figure 2.
Figure 2: Fluorescein Standard curves with their respective linear equations. (See normal scale left and log scale right).
Cell measurement
Transformations: Due to execute the best Cell Measurement, we performed different attempts using different transformation protocols in order to compare those and then use the best protocol.
Attempt 1: In this experiment we used electro-competent cells prepared in our laboratory, on the other hand we resuspendended and used the amounts specified in http://parts.igem.org/Help:Protocols/Transformation → the transformation protocol recomended by iGEM. This attempt failed because just a few colonies grew, according to the analysis of the controls used and the quantity of colonies that grew we decided to increase the amount of DNA in the next attempts.