Reliable and repeatable measurement which is a key component to synthetic biology is the aim of the iGEM InterLab Study. We took part in the 5th international interLab measurement study and the core question is:
“Can we reduce lab-to-lab variability in fluorescent measurements by normalizing to absolute cell count or colony-forming units (CFUs) instead of OD? ”
Parts
Interlab test devices used, as set by the iGEM measurement committee
Device | Part Number | Location |
---|---|---|
Negative control | BBa_R0040 | Well 2D |
Positive control | BBa_I20270 | Well 2B |
Test Device 1 | BBa_J364000 | Well 2F |
Test Device 2 | BBa_J364001 | Well 2H |
Test Device 3 | BBa_J364002 | Well 2J |
Test Device 4 | BBa_J364007 | Well 2L |
Test Device 5 | BBa_J364008 | Well 2N |
Test Device 6 | BBa_J364009 | Well 2P |
Prepare competent cells (Escherichia coli strain DH5α) and perform transformation according to the cell measurement protocol. Meanwhile, we followed the calibration protocol supplied by the iGEM HQ to generate OD600 reference point, particle standard curve and fluorescence standard curve to calibrate the plate reader (96 well plate) and then measured both the fluorescence and absorbance of our samples in 0-hour and 6-hour.
We used the setted protocol that we obtained so as to ensure everyone was doing correctly and repeatable. These can be found at this page.
Results and discussion
We submitted the iGEM’s completed spreadsheet to the measurement committee before the deadline, and received confirmation that the results were shown here: 2018/7/30
OD600 calibration
LUDOX CL-X H2O | ||
---|---|---|
Replicate 1 | 0.059 | 0.039 |
Replicate 2 | 0.060 | 0.035 |
Replicate 3 | 0.060 | 0.035 |
Replicate 4 | 0.061 | 0.040 |
Arith. Mean | 0.059 | 0.039 |
Corrected Abs600 | 0.023 | |
Reference OD600 | 0.063 | |
OD600/Abs600 | 2.769 |
Fluorescein standard curves
Particle standard curve
Our goal is to reduce lab-to-lab variations in fluorescent intensity measurements, in order to test the variation in our sample, we caculated coefficient of variation for both colony, and compared the CV value by using different normalization method (fluorescence per OD or fluorescence per particle)
0th
6th
After 6 hr growth, we found that the variation of each type of cells (devices) is decreased, but the CV value of devices 4 and device5 are much higher than the others. The variation of fluorescence per OD is much smaller than fluorescence per particle. Furthermore, we also found that the data from colony 2’s is usually more steady compared with colony 1. To conveniently compare the normalized value with original value, we removed the data of devices 4 and device 5, and only compare to the 6-hour’s data of colony 2 .
According to the graph, we found that fluorescence per particle could decrease the variation but it depending on the devices. It seemed that usage ofe OD value to normalize is still better than particle. But one thing need to be pointed out that too steady normalized value will lose sensitivity when test the influences of some factors. You can see this excel file to check our raw and manipulated data.