Team:Munich/InterLab

Phactory

Interlab Study

We were amazed by the possibilities of the iGEM InterLab study project. This project is the biggest interlaboratory study ever done in synthetic biology. We know this is going to push the frontiers of science towards achieving true experimental reproducibility and systematic consistency.

Since these elements are also key-aspects of a manufacturing project such as Phactory, we took this opportunity to participate very serious.

Materials and Methods

Plate Reader: FLUOstar (BMG LABTECH)
Plate Reader Plates: 4titude Vision Plate 96 Well (black plate with clear flat bottom, rounded wells)
Calibration Material: LUDOX for absorbance and Fluorescein for fluorescence (provided in the IGEM distribution kit)
Flow Cytometer: Cube 8 Sysmex Partec
Calibration Particles: SpheroTech Rainbow calibration particles type RCP-30-5A
Microorganism: Escherichia coli DH5⍺ strains
Negative control:
Positive control:
Device 1:
Device 2:
Device 3:
Device 4:
Device 5:
Device 6:

We followed the plate reader protocol and the the flow cytometry protocol provided by iGEM HQ.

Plate Reader Configurations

Gain: 600
Number of flashes per well (absorbance): 20
Number of flashes per well (fluorescence): 20
Temperature: 28,9 °C
Excitation wavelength: 485 nm
Emission wavelength: 520 nm
Shaking: 500 rpm
Path length correction: off

Results and Discussion

A generic square placeholder image with rounded corners in a figure.
Fluorescence calibration was done by measuring Fluorescein at different concentrations.
A generic square placeholder image with rounded corners in a figure.
Particle calibration was done by measuring calibration particles at different concentrations.
A generic square placeholder image with rounded corners in a figure.
Absorbance (OD = 600 nm) mean values at the start and end of the experiment for all devices.
A generic square placeholder image with rounded corners in a figure.
Fluorescence (480/520 nm) mean values at the start and end of the experiment for all devices.
A generic square placeholder image with rounded corners in a figure.
Fluorescence normalized by particle also supports those results

Conclusion

Device 2 showed the best fluorescence results, even better than the positive control. Device 4 was the second one with the highest emission. Device 4 also emitted fluorescence comparable to the positive control. Devices 3 and 5 showed no fluorescence. Device 1 showed no fluorescence but was reported to be a difficult plasmid before

[compare iGEM Munich 2017].

Flow Cytometer Configurations

Excitation Wavelength: 488 nm
Trigger: side scatter

Results and Discussion

Data were collected for all devices at the start (0h) and the end (6h) of the experiment. The bacteria population can be differentiated from background events.

A generic square placeholder image with rounded corners in a figure.
Exemplary data: forward vs. side scatter plot for blank, Device 2 and Device 5. The gating excludes background events for the further analysis.
A generic square placeholder image with rounded corners in a figure.
Exemplary data: fluorescence histogram for Row A at the start point (0h).
A generic square placeholder image with rounded corners in a figure.
Exemplary data: fluorescence histogram for Row A after 6h.

For the calibration particles RCP-30-5A a histogram with at least 7 peaks could be recorded. The particles did not show a discrete population in the forward vs. side scatter plot, indicating an aggregation of the particles.

A generic square placeholder image with rounded corners in a figure.
Exemplary data: fluorescence histogram for the calibration particles.

Conclusion

No signals were detected for the negative control, Device 1 and Device 3 in rows A to D and the negative control, Device 3 and Device 4 in rows E to H respectively. This indicates that Device 1 and Device 4 were swaped accidentally in this rows. After 6h Device 4 and Device 5 displayed in the fluorescent histograms a mixed population.