One of the biggest problems in synthetic biology is that there is a big variability in the results you get from different labs. The goal of the interlab study is to minimize this variability, so that the measurements are both reliable and repeatable for labs all over the world. This year’s interlab study focus on the question: Can we reduce lab-to-lab variability in fluorescence measurements by normalizing to absolute cell count or colony-forming units (CFUs) instead of OD? To answer this question two approaches were used [1]:
- Converting between absorbance of cells to absorbance of a known concentration of beads.
- Counting colony forming units from the sample.
Calibration
There were made three sets of calibration measurements: OD600 reference point, particle standard curve and fluorescein standard curve.
The first calibration measurement was to get a conversion factor so that we could easily convert absorbance (Abs600) to OD600. Here we used the LUDOX CL-X. The results were:
LUDOX CL-X | H2O | |
---|---|---|
Replicate 1 | 0.057 | 0.038 |
Replicate 2 | 0.058 | 0.038 |
Replicate 3 | 0.056 | 0.038 |
Replicate 4 | 0.056 | 0.038 |
Arith. Mean | 0.057 | 0.038 |
Corrected Abs600 | 0.019 | |
Reference OD600 | 0.063 | |
OD600/Abs600 | 3.338 |
For the particle standard curve, a dilutions series of monodisperse silica microspheres was prepared and the absorbance was measured in a Tecan Infinite M200 Pro plate reader. This curve was used to convert absorbance to number of cells. Figure 1 shows the linear particle standard curve, and a mathematical relationship between absorbance and number of particles was estimated:
y = 3*10-9x + 0.0401
with a value of R² = 0.9995. A particle standard curve in log scale was also plotted as shown in Figure 2.
A dilution series for the fluorescein standard curve was also made. The fluorescence was measured in a Tecan Infinite M200 Pro plate reader with an excitation wavelength 494nm and an emission wavelength 525nm. Figure 3 shows the relationship between fluorescence and the concentration of fluorescein. A plot in logarithmic scale was also made (Figure 4).
Cell measurements
In this study, we transformed Escherichia coli DH5α with the following plasmids from the iGEM distribution kit:
Device | Part number | Plate | Location |
---|---|---|---|
Negative control | BBa_R0040 | Kit Plate 7 | Well 2D |
Positive control | BBa_I20270 | Kit Plate 7 | Well 2B |
Test Device 1 | BBa_J364000 | Kit Plate 7 | Well 2F |
Test Device 2 | BBa_J364001 | Kit Plate 7 | Well 2H |
Test Device 3 | BBa_J364002 | Kit Plate 7 | Well 2J |
Test Device 4 | BBa_J364007 | Kit Plate 7 | Well 2L |
Test Device 5 | BBa_J364008 | Kit Plate 7 | Well 2N |
Test Device 6 | BBa_J364009 | Kit Plate 7 | Well 2P |
The day after, two colonies from each plate were inoculated in LB-medium + Chloramphenicol. The cells grew overnight. OD600 was measured and the cultures were diluted to a target OD600 of 0.02 in a final volume of 12 mL. Abs600 and fluorescence were measured after 0h and 6h incubation with four replicates.
Net Abs600
Figure 5 and Figure 6 show the Abs600 measured at 0 h and after 6 h of incubation from cell colony 1 and cell colony 2, respectively. By comparing the absorbance between the two colonies after 6 h incubation, the cells growth with different rate. However, for both cell colonies, the lowest net absorbance was detected from cells with device 1 or device 5 compared with the cells that did not carry either of these two devices.
Net Fluorescein a.u
Figure 7 and Figure 8 show the fluorescence of the different devices at 0h and after 6h incubation for colony 1 and colony 2, respectively. The highest net fluorescein a.u was detected from cells which carried device 4, while cells with device 3 had the lowest net fluorescein a.u (Figure 7 and Figure 8). However, the net fluorescein a.u measured from colony 1 tend to be higher, compared with colony 2.
µM Fluorescein/OD
The highest µM Fluorescein/OD was detected from cells which carried device 1, 4 or 5. On the other hand, cells which carried device 3 had the lowest µM Fluorescein/OD, which is comparable with the negative control. Figure 9 and Figure 10 show the µM Fluorescein/OD at 0 h and after 6 h incubation of cell cultures from colony 1 and colony 2, respectively.
Molecules of equivalent fluorescence level per particle (MEFL/ particle)
Like the µM Fluorescein/OD, device 1, 4 and 5 had the highest MEFL/ particle value, while device 3 had the lowest MEFL/ particle value (Figure 11 and Figure 12).
Colony forming units
The second approach of this interlab study was to find colony forming units (CFUs) per 0.1 OD600 E. Coli cultures. Starting samples were prepared from the positive and negative control by diluting the overnight culture 20 times until OD=0.1 for the cell culture minus the blank measurement. Dilution series was prepared and 36 plates were aseptically spread with dilution series 3, 4 and 5.
Given that the cultures were diluted to target OD600 = 0,1. The Abs600 = 0,0678 (including blank) was obtained by using the factor OD600/ Abs600 = 3,338 to convert OD600 to Abs600. According to the standard particle curve, the predicted number of cells will be 0,93*107. By comparing the values of CFU/mL from Table 5 with the predicted number of cells (0,93*107), these values have a relatively large variance, and a large deviance from the predicted number of cells.
Dilution 3 | Dilution 4 | Dilution 5 | |
---|---|---|---|
Final dillution factor | 8*104 | 8*105 | 8*106 |
The plates were incubated at 37°C overnight. The day after, the colonies were counted and the results are given in table 4 and 5:
Positive control | Negative control | |
---|---|---|
Culture 1, dilution 3 | 194.3 | 117.3 |
Culture 1, dilution 4 | 7.7 | 31.0 |
Culture 1, dilution 5 | 1.3 | 2.3 |
Culture 2, dilution 3 | 104.0 | 87.0 |
Culture 2, dilution 4 | 18.7 | 5.7 |
Culture 2, dilution 5 | 1.3 | 0.7 |
Positive control [CFU/mL] | Negative control [CFU/mL] | |
---|---|---|
Culture 1, dilution 3 | 1.6*107 | 0.9*107 |
Culture 1, dilution 4 | 0.6*107 | 2.5*107 |
Culture 1, dilution 5 | 1.1*107 | 1.9*107 |
Culture 2, dilution 3 | 0.8*107 | 0.7*107 |
Culture 2, dilution 4 | 1.5*107 | 0.5*107 |
Culture 2, dilution 5 | 1.1*107 | 0.5*107 |
Conclusion
The results obtained from this experiment, indicate that the two approaches for reducing the lab-to-lab variability in fluorescence measurement, do not seem to be reliable. Neither method 1 nor method 2 gave results that could return to the cell concentration in the sample, due to a relatively large variance and deviation.
Sources: [1] https://2018.igem.org/Measurement/InterLab