Calibration 3: Fluorescence Standard Curve - Fluorescein Protocol
Fluorescein was used to creat a standard curve for fluorescence as using GFP protein with all of the participating teams would be expensive. Figure 3 and Figure 4 show our results.
Figure 3. Fluorescein standard curve created by measuring Fluorescence of a series dilution of fluorescein.
Figure 4. Fluorescein standard curve created by measuring Fluorescence of a series dilution of fluorescein.
Results and Discussion
Colony forming unit per 0.1 OD600 E. coli cultures
Replicates of a starting overnight cultures were all diluted to OD600 and then serial diluted and plated. CFUs (colony forming units) were counted and assuming 1 bacterial cell will give rise to 1 colony, number of CFUs should directly correlate to concentration of cell culture. Figure 5 show our results.
Figure 5. Estimated CFU in a starting 1 ml sample of OD600 = 0.1. 1.1-4.3 are replicates of BBa_R0040 and BBa_I20270 which are the positive (1.1-2.3) and negative controls (3.1-4.3) respectively. CFU/mL was calculated by dividing the colonies formed on plates by its dilution factor. D4 is a dilution factor of 1/(8x105,) and D5 is a dilution factor of 1/(8x106).
Figure 6. Fluorescence measurements of colonies taken at 0 hours and 6 hours. Colonies show wide variability of fluorescence with notably colony 2 of negative control showing fluorescence likely suggesting there was have been contamination from one of the other colonies or the Edinburgh_UG team.
Figure 7. Absorbance600 measurements of colonies taken at 0 hours and 6 hours. Colonies show a range of 0D600 from 0.2 to 0.37. However there does not seem to be any link between OD600 at 0 hours or at 6 hours.
Conclusion
Looking at our data, we can see that our data suggest that the amount of cells in a 1 mL culture is between 60,000,000 and 100,000,000. Our Dilution 4 data estimates the number of cells to be much lower than our dilution 5 data almost by half.